The software events and methods have been modified to increase sensitivity and accuracy for TOC, TC and TN methodologies over all ranges. In addition, Tekmar's applications laboratory has determined that 1-2% Hydrochloric Acid (HCl) acid should be used for the removal of Inorganic Carbon (IC) when using TOC/TN methods. Use of 20% phosphoric acid may yield decreased recovery on some nitrogen compounds if used with TOC/TN methods. The life of the corrosive scrubber should not be significantly impacted by this change. However, the corrosive scrubber (copper and tin granules) should continue to be checked for use on a daily basis. When the copper granules begin to dual in color and brightness, they should be replaced.
Previous changes:
Note: All previous changes are included within the latest Release
- 4.5 Service Pack 3b (SP3(b)):
Pumper Module Communication Update
* This update slows the rate of query status requests from the pumper module in order to provide a wider safety margin for the pumper communication tolerance. This is helpful on newer pumpers.
- 4.5 Service Pack 2 (SP2):
Service Pack 2 Methods
* The Injection Volume on the cleaning procedure went from 0 to 0.5 ml. This eliminated a “divide by zero” error.
Service Pack 2 Events
* The furnace gas is now left on throughout the analysis. This helps to eliminate backpressure effect during bake line acquisition.
- 4.5 Service Pack 1 (SP1):
Method Editor
* The Method Editor would at times draw the incorrect version when the user selected to “View Previous Version”. This has been corrected.
* The rules that restricted users from entering certain values in a user method have been relaxed somewhat to allow the more flexibility.
* The Method Editor had some trouble when dealing with method names that were one character in length. This has been corrected.
Events 1.5
* The version of the Event data files has been updated to version 1.5. This latest change involves removing additional water when running the Condition Catalyst Method.
|